Improved Method for Determination of Raspberry Ketone in Fragrance Mist by HPLC-Fluorescence Analysis after Pre-Column Derivatization with 4-(N,N-Dimethylaminosulfonyl)-7-(N-chloroformylmethyl-N-methylamino)-2,1,3-benzoxadiazole ()
1. Introduction
Raspberry ketone {RK, 4-(4-hydroxyphenyl)butan-2-one}, which is present in red raspberry, was reported to have an anti-obesity effect [1] [2] [3] , to possess a depigmenting activity in C57 black mice [4] , to inhibit melanogenesis in cultivated murine B16 melanoma cells [5] and to suppress proliferation of normal melanocytes [6] . These findings indicated that RK would be useful for skin whitening, and RK-added cosmetics are already commercially available. In addition, cosmetics containing rhododendrol {4-(4-hydroxyphenyl)-2-butanol}, which structurally resembles RK, were withdrawn from the market in July 2013, because of observation of leukoderma on consumers’ hands and face [7] . Thus, it seems important to measure the RK level in cosmetics, since RK may induce advertised effects like those of 4-(4-hydroxyphenyl)-2-butanol.
High-performance liquid chromatographic (HPLC) methods with ultraviolet-visible absorption detection and gas chromatographic (GC) method using a flame ionization detector have been developed to determine raspberry ketone [8] [9] [10] . Although a coupled loop-type HPLC-GC method was established for analysis of RK in raspberry sauce, this technique may be unsuitable for broad application [11] . In addition, GC and HPLC methods coupled with mass spectrometry have been utilized for RK determination, but the equipment is expensive and complex [12] [13] . Very recently, HPLC-fluorescence determination of RK was developed using pre-column derivatization with 4-hydrazino-7-nitro-2,1,3-benzoxadiazole (NBD-H) [14] . While the assay possessed low cost and exhibited a high sensitivity (absolute amount of 1.8 pmol), optimum derivative conditions were relatively severe (80˚C, 20 min).
RK possesses a phenolic hydroxyl group as well as a carbonyl group in the chemical structure. NBD-H reacts toward a carbonyl group of RK. Several papers about HPLC-fluorescence method after mild and rapid derivatization (at room temperature, for 3min) using acyl chloride compounds toward a phenolic hydroxyl group were reported [15] [16] [17] . Therefore, in the present study, an HPLC-fluorescence method for determination of RK in a fragrance mist after pre-column derivatization with an acyl chloride compound, 4-(N,N-dimethylaminosulfonyl)-7-(N-chloroformylmethyl-N-methylamino)-2,1,3-benzoxadiazole (DBD-COCl), was established. The derivatization scheme is shown in Figure 1.
![]()
Figure 1. Derivatization reaction scheme for RK.
2. Experimental
2.1. Materials
RK and DBD-COCl were purchased from Tokyo Chemical Industry Co., Ltd. (Tokyo, Japan). Fragrance mist was obtained via the internet. Acetonitrile and distilled water for HPLC was purchased from Kanto Chemical Co., Inc. (Tokyo, Japan). Trifluoroacetic acid was obtained from Wako Pure Chemical Industries (Osaka, Japan).
2.2. HPLC Instrumentation and Chromatographic Conditions
The HPLC system consisted of a model L-6200 pump (Hitachi, Tokyo, Japan), a Rheodyne injection valve (Cotati, CA, U.S.A.) with a 20-μL loop, and a model RF-10AXL fluorescence spectrophotometer (Shimadzu, Kyoto, Japan) with excitation at 440 nm and emission at 543 nm. A 150 mm × 3.0 mm i.d. HPLC column (ODS-4, GL Science, Tokyo) containing 5 μm particles of C18 packing material was used. Peak quantification was performed using a model C-R3A Integrator (Shimadzu). The mobile phase was prepared by the addition of acetonitrile (450 mL) to 550 mL of Milli-Q water containing trifluoroacetic acid (0.1 %v/v). Samples were eluted from the column at room temperature at a flow rate of 0.5 mL/min.
2.3. Preparation of Standard Solutions, Lower Limits of Detection and Quantification
After dissolving RK (5.0 mg) in methanol (1.0 mL), standard stock solution (50 μg/mL) was prepared by dilution with distilled water and stored at 4˚C. Working standard solutions (0, 0.05, 0.1, 0.25, 0.5, 1, .1.5 and 2.5 μg/mL) were prepared by dilution with distilled water. Determination of lower limit of detection (LOD) and lower limit of quantification (LOQ) was based on the standard deviation (SD) of the response and on the slope of the calibration curve (0.05, 0.1, 0.25 μg/mL), according to the following equations: LOD = 3.3 × SD/slope, LOQ = 10 × SD/slope.
2.4. Derivatization
Borate buffer (0.1 M) was adjusted to pH 9.0 by the addition of NaOH. Borate buffer (50 μL) was added to diluted standard samples (50 μL). DBD-COCl solution in acetonitrile (2 mg/mL, 100 μL) was added and vortexed. The mixture was allowed to react for 3 min at room temperature. Saturated L-aspartate solution (50 μL) was added to stop the reaction, and aliquots (20 μL) were injected into the HPLC system.
2.5. Application to Fragrance Mist Sample and Evaluation of Recovery
Fragrance mist was diluted 1600-fold with distilled water, and then analyzed as described above. Fragrance mist (1.0 mL) was spiked with 1.0 mg of RK, and spiked sample was diluted 1600-fold with distilled water. The resulting sample was analyzed to determine recovery of the added standard in order to assess the accuracy of the method.
3. Results and Discussion
3.1. Time Course of Derivatization of RK with DBD-COCl
For the time course study, the reaction time was set at 1.5, 3, 5, 10, and 15 min. RK (50 μL, 1 μg/mL), borate buffer (50 μL, pH 9.0) and DBD-COCl (50 μL, 2 mg/mL) were mixed as described in 2.4. Derivatization. The maximum peak area was reached at 3 min (Figure 2(a)). Next, pH dependency (pH 7.5 to 10.0) was examined at the derivatization time of 3 min (Figure 2(b)). Peak area of DBD-CO-RK
(a)
(b)
Figure 2. (a) Plot of peak area versus reaction time for formation of the RK derivative with DBD-COCl. Data are mean values of two experiments; (b) Plot of peak area versus various pH values for formation of the RK derivative with DBD-COCl. Data are mean values of two experiments.
at pH 9.0 showed the maximum. The peak area was remarkably low at neutral pH, and was tended to decrease at pH 9.5 and 10.0. Thus, the derivatization time of 3 min at pH 9.0 was chosen for the assay.
3.2. Chromatogram
Figure 3 shows typical chromatograms obtained from (A) standard sample (1 μg/mL) and (B) test sample of a fragrance mist (diluted 1600-fold with distilled water). The retention time of DBD-CO-RK derivative was 16.6 min. A peak of DBD-CO-RK derivative was observed in the test sample. The running time was set at 30 min, because unknown peak was detected at 28.8 min.
3.3. Standard Curves of RK
The standard curve of RK was constructed by plotting integrated peak area vs. concentration. The calibration plot was linear (slope, 903.5; intercept, +36.58) in the range of 0.05 to 2.5 μg/mL with a correlation coefficient (r2) value of 0.9988. The lower limit of detection for RK was estimated as described above. The values of the lower limit of detection and quantification were 0.01 μg/mL (absolute amount of 0.3 pmol, 0.05 ng) and 0.03 μg/mL (absolute amount of 0.9 pmol, 0.15
![]()
Figure 3. Typical chromatograms obtained for (a) standard sample (1 μg/mL) and (b) test sample of a fragrance mist (1600-fold diluted sample). Retention time of DBD-CO-RK derivative (arrowed peak): 16.6 min.
ng), respectively.
RK is a volatile compound, and has been analyzed by GC-mass spectrometry [10] and GC with a flame ionization detector [11] , but those reports did not clearly describe the sensitivity of their methods. Also, the lower limit of detection of RK with the previous HPLC methods was not established [8] [9] [13] . However, HPLC analysis using a pre-column derivatization with NBD-H showed detection limits of 1.8 pmol [14] . Therefore, the sensitivity of the present method (0.3 pmol RK) appears to be improved.
3.4. Precision and Accuracy
Precision and accuracy in intra-day and inter-day assays of RK are shown in Table 1. In the intra-day assay, the range of standard deviation was within 4.3% to 8.6% of the mean and recovery was within the range of 96.8% to 102.2%. In the inter-day assay, the range of standard deviation was within 5.7% to 10.0% of the mean and recovery was within the range of 97.6% to 100.8%.
3.6. Analysis of a Fragrance Mist
The developed method was used to determine RK in fragrance mist and in fragrance mist spiked with authentic standard. The concentration of RK in fragrance mist was 1.20 ± 0.08 mg/mL (average ± S.D., range, 1.10 to 1.31 mg/mL, n = 5, data not shown). Recovery of RK from spiked fragrance mist was 91.8 ± 5.4% (average ± S.D., range, 84.2 to 98.2%, n = 5, data not shown). The present concentration of RK in fragrance mist was almost fully consisted with previous data (1.18 ± 0.07 mg/mL) [14] . Those data suggest the present assay would be suitable for RK determination.
4. Conclusion
An improved HPLC-fluorescence method for determination of RK in fragrance mist has been developed by using DBD-COCl as a fluorescence-labeling reagent. Since derivatization conditions are mild and rapid, the present method would be more useful than previous method [14] . Also, sensitivity of the method is 6-fold improved. The mean content of RK in the mist (1.00 mL) was found to be 1.20 ±
![]()
Table 1. Intra- and inter-day assay reproducibility for determination of RK.
0.08 mg. This system should be suitable for routine quality assessment of fragrance mist and should be readily adaptable for measurement of RK levels in other cosmetics.