Comparative Study between Field-in-Field and IMRT Techniques in Prostate Cancer Radiotherapy: A Treatment Planning Study ()
Received 27 October 2015; accepted 1 February 2016; published 4 February 2016

1. Introduction
Three-dimensional (3D) conformal radiation therapy has been demonstrated to achieve improvement of tumor targeting and to reduce normal tissue volume exposed in several malignancies [1] [2] .
To improve dose distribution, the FIF technique has been used in the treatment of certain cancers. It is a manually based forward intensity-modulated radiotherapy (forward-IMRT) plan for which the calculated dose is adjusted in certain dose distribution areas by designing multiple lower-weighted reduction fields based on the main field [3] -[5] .
Intensity-Modulated Radiation Therapy (IMRT) is a radiation therapy technique in which non-uniform fluence is delivered to the patient from any given direction of the treatment beam to improve the composite dose distribution. The treatment criteria for plan optimization are determined by the planner and the optimal intensity profiles for a given set of beam directions are specified through “inverse planning” [6] [7] . The intensity files thus generated are electronically delivered to the linear accelerator (Linac), which is computer controlled, equipped with the required hardware and software to deliver the calculated intensity-modulated beams (IMBs) [8] -[10] .
2. Materials and Methods
2.1. Acquisition and Simulation
Ten prostate cancer patients underwent a Computed Tomography (CT) scan with 2 mm slice thickness. All sets of CT cuts were transferred to Focal system. Tumor volumes such as gross target volume (GTV), clinical target volume (CTV), and planning target volume (PTV) [11] as well as Organs at Risk (OARs) had been delineated by the physician on each axial slice, and then CT slices were transferred to the treatment planning system (TPS).
2.2. Treatment Planning System
Three-dimensional treatment planning system (TPS) Computerized Medical Systems (CMS) XiO software (release 4.64) was used to carry out dose calculation for all patients under study by super position algorithm. The TPS has both inverse and forward planning facilities. In inverse planning: after the beams are set, the inverse IMRT planning facility was used to make an automatic segmentation of all beams using the Multileaf Collimator (MLC). In forward IMRT planning (Field-in-Field (FIF)), the beam parameters were manually adjusted in a try and error process to achieve a good dose distribution to the PTV and at the same time to spare the OARs from receiving high doses.
2.3. Beams Arrangement
The beam arrangement was determined by the size and location of the tumor. Major seven fields with equal values of dose were used to plan each patient. In inverse planning, the beams are spread around the target with equispace and to avoid the opposing fields an odd numbers of the treatment fields were used. IMRT dose constraints for both target and OARs were entered to Xio TPS [12] - [14] . Also some conditions were given to the inverse TPS which were the minimum, goal and maximum radiation doses for the target volume, the power and weight of each structure. Several trials were taken place by Inverse TPS to achieve the IMRT dose constraints and a homogeneous dose distribution [10] [12] . Then FIF technique is used for the same selected patients where some of the manually set beams were manually segmented and one or more of sub-fields were set inside the main field(s) with small doses aiming to get rid of over doses in the main field(s) and therefore to achieve a homogeneous dose distribution.
2.4. Evaluation of the Treatment Plans
The treatment plans can be qualitatively and quantitatively evaluated by many tools which are already included into Xio TPS. In our study, we used two main treatment plan evaluation tools; the visual slice-by slice review of the treatment plans using isodose lines distribution as a qualitative evaluation for the treatment plans which is important to know the location of the hot and cold areas and review dose distribution to both of the clinical target and OARs and the Dose Volume Histogram (DVH) was generated to evaluate the dose to the different structures in different treatment plans and it can be used as a quantitative evaluation for the treatment plans [15] - [17] .
3. Results
3.1. FIF versus IMRT Radiotherapy Planning Techniques in Prostate Tumors
3.1.1. According to a Qualitative Evaluation Tool (the Axial Cuts and DRRs)
The next Figure 1 showed colored print screens of two axial cuts and DRRs for every one of ten prostate cancer patients planned using FIF (Group (a)) and IMRT (Group (b)) radiotherapy treatment techniques. Figure 1(a) and Figure 1(b) showed the dose distribution to the treatment target (PTV), the beams arrangements, some isodose lines and how the OARs are protected from being exposed to high radiation doses in each one of the FIF and IMRT techniques. As a qualitative evaluation; from the next figure, it is noted that the IMRT achieved a better dose distribution to target of treatment (PTV) than that dose distribution achieved by the FIF. But the FIF achieved a better OARs radiation dose sparing than that achieved by the IMRT.
3.1.2. According to a Quantitative Evaluation Tool (from the DVHs)
1) For the PTV:
Table 1 showed the Mean and Standard Deviation (±SD) of volume % of PTV covered with 107% of the total prescribed dose (PTV 107%), PTV 95%, PTV 90%, PTV 70%, dose % received by 98% of PTV volume (D 98%), D 95%, D 50% and D 2% in Method 1 (FIF) and Method 2 (Inverse-IMRT) respectively. It is noticed that the Inverse-IMRT achieved higher mean values for the all mentioned PTV related results and achieved lower SD values for the same PTV related results than FIF. Figure 2 showed the Mean and SD for the Homogeneity Index (HI) of the all above PTV related results in all patients under the study in Method 1 and Method 2, and they are (0.197 ± 0.17, 0.147 ± 0.07) respectively, from which we noted that there is no significant difference between the two methods. Also Figure 3, Figure 4 showed the Mean and SD of Dose Global Max and Mean Dose to PTV in Method 1 and Method 2 which are (98.1 ± 1.75, 100.8 ± 0.45), (101.5 ± 0.97, 104.6 ± 2.54) respectively, where the all Mean values for the two Methods are within the acceptable dose range to the PTV. It is noted that there are some big SD values in Method 1 (FIF) which are due to the resulted heterogeneous dose distribution to the PTV in FIF because of being it so closed to OARs.
2) For the OARs:
The next Table 2 showed Mean and Standard Deviation (±SD) of Organs at Risk (Rectum, Bladder and Right and Left Femoral Heads) dose parameters for both Inverse-IMRT and FIF techniques respectively in all cases under the study. It is noted that FIF achieved a better sparing of both Rectum and Bladder from receiving high doses than IMRT and. But for the Right and Left Femoral Heads, the IMRT achieved better protection than FIF.
4. Discussion
Being of many international publications studied the use of Inverse-IMRT technique in prostate cancer radiation
![]()
Table 1. Mean and Standard Deviation (±SD) of PTV dose parameters for both FIF and Inverse-IMRT.
![]()
![]()
(a) (b)
Figure 1. Group (a) and Group (b): Show two axial cuts and DRRs for every one of ten prostate cancer patients showing the beams arrangement, the dose distribution covering the PTV and protection of OARs in FIF (Group (a)) and IMRT (Group (b)) radiotherapy treatment plans.
![]()
Figure 2. Mean and Standard Deviation (±SD) Homoge- neity Index (HI) for PTV for IMRT and FIF techniques in ten patients with Prostate tumors. Calculation formula: HI = (Dose Max. − Dose Min.)/Dose Mean in PTV.
![]()
Figure 3. Mean and Standard Deviation for Mean Dose % Received by PTV for IMRT and FIF techniques in ten pa- tients with Prostate tumors.
![]()
Figure 4. Mean and Standard Deviation for Dose Global Maximum % Received by PTV for IMRT and FIF tech- niques in ten patients with Prostate tumors.
![]()
Table 2. Mean and Standard Deviation (±SD) of Organs at Risk dose parameters for both Inverse-IMRT and FIF.
therapy, the lack of the publications that discussed the use of the Field-in-Field (FIF) technique (which is a form of the Forward-IMRT) in the prostate cancer radiotherapy was the main reason of choosing the point of this study. Also it is worth to be mentioned that we noted that many centers like our centers at which we are work, still use the FIF more than they use the IMRT in prostate cancer radiotherapy.
5. Conclusion
According to the all previous results, we conclude that the IMRT planning technique achieved better dose coverage to the PTV than the FIF planning technique but except RT and LT Femoral Heads, FIF achieved a better protection to the Rectum and the Bladder (OARs) than IMRT. So we conclude and recommend that the IMRT technique is better in the prostate cancer radiotherapy than the FIF technique [18] - [22] .
Limitations
There are no limitations regarding to the application of this study for treating the prostate cancer patients as long as the Inverse-IMRT planning was available.
Acknowledgements
The authors thank Prof. Dr. Assem Rostom, consultant clinical oncologist, the general manager of Alexandria Ayadi Almostakbl Oncology Centre (AAAOC) and Dr. Mostafa Ali Elnaggar, the medical manager of Alexandria Ayadi Almostakbl Oncology Centre (AAAOC), for their continuous and valuable support and encouragement.
Abbreviations and Acronyms
IMRT: Intensity-Modulated Radiation Therapy
AAAOC: Alexandria Ayadi Almostakbl Oncology Center
FIF: Field-in-Field
PTV: Planning Target Volume
OARs: Organs at Risk
DVHs: Dose Volume Histograms
3D: Three-dimensional
Linac: Linear accelerator
IBMs: Intensity-Modulated Beams
CT: Computed Tomography
GTV: Gross Target Volume
CTV: Clinical Target Volume
TPS: Treatment Planning System
CMS: Computerized Medical System
XiO: Name of three dimensions treatment planning system
SD: Standard Deviation
HI: Homogeneity Index
NOTES
![]()
*Corresponding author.