Radiation Risks of Kidney during Abdominal CT Procedures in Morocco: A Multicentre Study ()
1. Introduction
Medical imaging is a valuable and indispensable tool for the detection, diagnosis and efficient control of many diseases. Every year, approximately 3.6 billion diagnostic radiology procedures are performed worldwide [1]. Exposure to X-rays during imaging procedures can generate biological effects. Ionizing radiation can cause cell death by apoptosis or failure of induced reproduction, which can lead to changes in genes involved in the regulation of cell growth, loss of normal nuclear structure, DNA degradation and tumorigenesis [2]. These effects include cancer and hereditary effects, which increase an individual’s lifetime risk of developing cancer or hereditary effects in future generations. Tissue response effects have specific threshold doses inducing radiological risks at relatively high doses [3]. The International Commission on Radiological Protection (ICRP) has adjusted the nominal radiological detriment coefficients for cancer and hereditary effects as follows: 5.5 × 10−2 and 0.2 × 10−2 Sv−1 for the whole population [3]. In addition to these effects, radiation exposure is associated with certain diseases (non-cancer effects), such as respiratory diseases, stroke, heart disease and digestive disorders [3] [4]. Although the radiological risks of non-cancerous diseases at low doses remain uncertain, doses delivered should be kept at minimum values to ensure patient protection [2] [3].
Chronic kidney disease is a real public health problem worldwide, economically, medically, and socially. It affects about 3% of the Moroccan population [5]. The multi-detector scanner is the reference imaging modality for urothelial tumor assessment thanks to its excellent spatial and temporal resolution [6] [7]. It allows the analysis of the renal parenchyma at the different phases of its development as well as the secretion of contrast agents in the urinary cavities. As a result, patient care has been significantly improved by accurate diagnosis of urinary system disorders compared to other imaging modalities. Repeated exposure of the kidneys during certain CT scans increases the effective dose received by this organ. However, CT procedures present a radiogenic risk due to the high doses delivered during image acquisition, which depends on age, gender and health status [8]. Previous studies have estimated that 1.5% - 2% (29,000 cancer cases) of cancers diagnosed each year in the United States are associated with CT scan exposure [4] [9].
In Morocco, the scarcity of studies on exposure to ionizing radiation during medical examinations of the population [10] [11] and on the establishment of diagnostic reference levels [12] [13] [14] [15] [16] has led us to explore the exposure of the kidneys during CT examination of the abdomen.
The objective of this study was to evaluate the effective dose of the CT scan examination as well as the renal dose in order to estimate the likely biological effects of such exposure.
2. Materials and Methods
2.1. Data Collection
This is a prospective study that was conducted in three Moroccan hospitals. A sample of 120 adult patients referred for a CT scan of the abdomen was collected at a rate of 40 patients per radiology department. For each patient we collected these demographic data as well as exposure parameters (kV, mAs, Pitch, CTDIvol, DLP etc.) on the console of a Philips CT scanner with 2 slices for the hospital H1, a GE CT scanner with 16 slices for the hospital H2 and a Toshiba CT scanner with 64 slices for the hospital H3.
The effective dose and renal dose were estimated according to the recommendations of the International Commission on Radiological Protection [3]. The formulas below were used:
(1)
with:
- Deff: The effective dose received per patient for a given CT scan.
- DLP: Dose length product for a given scan examination.
(2)
with:
- Dkidney: Renaldose received per patient for a given CT scan.
- CTDIw: Computed Tomography Dose Index weighted.
- mAs: Load for a given CT scan.
2.2. Cancer and Hereditary Risks Assessment
The overall cancer risk per procedure was obtained by multiplying the effective dose per Sievert (Sv) by the coefficient 5.5 × 10−2 Sv−1 according to the ICRP 103 formalist [3].
The risk for hereditary diseases up to the second generation per procedure was estimated by multiplying the effective dose expressed in Sievert (Sv) by the coefficient 0.2 × 10−2 Sv−1 according to ICRP 103 [3].
3. Results
Data from 120 patients referred for a CT scan of the abdomen were collected from 40 patients per radiology department in the three hospitals involved in the study. Table 1 includes age, weight, exposure parameters (kV, mAs) as well as dosimetry parameters (CTDIvol, DLP) for the abdominal procedures performed.
The age of the patients included in this study varies between (41.00 ± 11.56) and (52.15 ± 19.38) with a mean value of (46.49 ± 14.16) years. For their weight, it varies between (71.44 ± 7.71) and (76.31 ± 6.84) with a mean value of (73.34 ±
Table 1. The mean and range values for Age, Weight, kVp, and mAs and for all the abdominal procedures and per Hospital.
7.58) kg. To study the effect of these variations on the dose results, the sample P-value correlation coefficient was calculated between weights, CTDIvol and DLPs for all examinations. It is 0.08 and <0.01 for weight vs CTDIvol and weight vs DLP, respectively.
For the exposure parameters, the voltage applied in the three hospitals is almost the same ≈ 120 kV and the current varies between 120 and 213 with an average value of 155 mAs. For dosimetric parameters, CTDIvol per sequence varies between 7.38 and 17.04 mGy with a mean value of 10.89 mGy and DLP per examination varies between 407.76 and 518.29 with a mean value of 438.73 mGy.cm.
The effective dose (E) received per patient during an abdominal CT scan varies between (6.29 ± 2.75) and (8.12 ± 1.83) mSv with a mean value of (6.67 ± 2.73) mSv. While the renal dose (DK) varies between (13.32 ± 5.15) and (26.99 ± 6.07) mSv with a mean value of (18.26 ± 7.74) mSv (Table 2). The distribution of these values according to the hospital variable shows a difference in mean effective dose of the order of 0.26, 0.38 and 1.45 mSv and a difference in mean renal dose of the order of 8.76, 4.94 and 0.48 mSv for H1, H2 and H3 respectively. However, the ANOVA test showed a significant difference between the three hospitals in terms of effective dose (F E (mSv) = 20.10; p ≤ 0.01) and not significant in terms of renal dose (F Dk (mSv) = 2.17; p = 0.119).
The effective dose E (mSv) and the renal dose Dk (mSv) received per patient during the CT scan of the abdomen can cause biological effects. The risk of induction of abdominal and kidney cancer is 3 to 4 and 7 to 13 per 105 procedures, respectively (Table 3). Kidney cancer per procedure is two to three times more likely than abdominal cancer. For the hereditary risk of abdominal and renal exposure, it is 12 to 16 and 27 to 54 per 106 procedures, respectively. The kidney stochastic effect per procedure is also two to three times more likely than that of the abdomen. However, the ANOVA test showed a non significant difference between the three hospitals in terms of carcinogenic risk (F kidney cancer risk = 1.97; p = 0.145; F CT cancer risk = 0.497; p = 0.609) and stochastic risk (F kidney hereditary risk = 1.382; p = 0.257; F CT hereditary risk = 1.211; p = 0.303).
4. Discussion
The average body weight of patients in the three hospitals included in the study ranged from (71.44 ± 7.71) kg to (76.31 ± 6.84) kg. This shows that the patients weighed more than 60 kg considered as average weight for Asian and African
Table 2. The mean, range, median, and for the effective dose E and kidney doseDK (mSv) for all procedures and per hospital.
Table 3. Cancer and hereditary risks per abdominal CT procedure and per hospital.
countries and 70 kg for developed countries [17]. This may also partly explain the slightly higher CT dose index (CTDI) values than those obtained by Shirazu et al. 2017 [18]. Differences in collected patient weights could influence the calculated effective doses due to the correlation found between weights and Dose Length Product (DLP) values (p < 0.01) and not renal doses (p = 0.08).
For exposure factors, the applied tube tension is almost the same for all three hospitals. It varies between (119.23 ± 8.62) and (120.00 ± 0.00) kVp with a mean value of (119.81 ± 4.15) kVp, corresponding to the trends seen in most previous studies [17] [19]. On the other hand, the current applied per procedure varies between (120.00 ± 0.00) and (212.81 ± 21.98) mAs with a mean value of (154.96 ± 41.83) mAs. These values are comparable to those reported by Sulieman et al. 2020 [19].
Analysis of abdominal image data from the three hospitals showed that the mean value of the CTDIvol was between (7.38 ± 2.85) and (17.04 ± 3.07) mGy with a mean value of (10.89 ± 4.77) mGy. The latter is almost twice as high as that reported by Shirazu et al. 2017 [18]. It is also higher than that recommended by the ICRP [3]. The mean DLP was (438.73 ± 179.93) mGy.cm with a minimum value recorded as (407.76 ± 181.29) mGy.cm and maximum value recorded as (518.29 ± 140.81) mGy.cm. It was lower than the value found by Shirazu et al. 2017 (935.5 mGy.cm) and also lower than that recommended by International Commission on Radiological Protection (ICRP) Publication 103 (780 mGy.cm) [3] [18]. In addition, about 7% of the DLPs were above the ICRP recommended doses [3].
The effective dose calculated from the DLP using the conversion factor of ICRP Publication 103 shows a variation from a minimum of (6.29 ± 2.75) mSv to a maximum of (8.12 ± 1.83) mSv with an average value of (6.67 ± 2.73) mSv. Renal dose was also estimated using ICRP Publication 103 as for effective dose. The calculated mean renal dose was (18.26 ± 7.74) mSv with (26.99 ± 6.07) mSv and (13.32 ± 5.15) mSv as the maximum and minimum values recorded at H2 and H1 hospital respectively. It should be noted that all renal doses obtained exceeded the ICRP and EC recommended value of 12 mSv [3]. However, the renal dose can be up to about three times that proposed by the ICRP and the EC [3]. This higher value can be explained by the broad parameters used by the scanning procedure pitch ≈ 1.75 and load = 154 mAs which were greater than 1.00/100 mAs and 0.813/75 mAs used by Salama et al. 2017 and Shirazu et al. 2017, respectively [17] [18]. For the effective dose, all our values are lower than the value obtained by Shirazu et al. 2017 and that recommended by the ICRP and the EC [3] [18].
5. Conclusion
The assessment of the effective and renal doses for 120 abdominal CT scan examinations in three different Moroccan hospitals allowed us to characterize the biological risks of exposure. The results obtained show a wide variation in exposure parameters and exposure doses during abdominal CT scans from one hospital to another. Even so, the average effective dose was generally lower than that recommended by the International Commission on Radiological Protection. This may not be sufficient, as some of the individual renal doses exceed the recommended values by up to 300%, which can lead to a consequent prognosis. We therefore recommend the establishment of the national diagnostic reference level for the CT scan examination of the abdomen and also the training of our clinicians on the radiation protection of the patient.
Acknowledgements
We express our gratitude and thanks to the staff of the radiology departments of the three hospitals for their help during collection of data.