Investigating the Fluid Seal of Supraglottic Airway Devices in Humans Using Indicator Dye via the Drainage Tube: A Potential Roadmap for Future Studies

Abstract

Background: The fluid seal of supraglottic airway devices (SGA) protects the airway from fluid contamination. We evaluated the suitability of indicator dye placement in the upper digestive tract of anesthetized patients combined with fiberoptical tracing to investigate the fluid seal of SGA. Methods: Patients swallowed a capsule of indigo carmine green (ICG) prior to induction of anaesthesia. After induction of anesthesia, one of two different SGA (either an i-GelTM or an LMA-SupremeTM (LMA-S)) was inserted after randomization. Methylene blue stained normal saline was injected through the proximal opening of drainage tube during mechanical ventilation as well as spontaneous breathing. We monitored regurgitation of ICG with a flexible fiberscope (FO) inserted through the drainage tube and checked for the appearance of methylene blue in the mask bowl with the FO inserted through the airway tube. Results: In thirty-six patients with an i-GelTM and 37 with a LMA-S no regurgitation of ICG was observed at the level of the upper oesophageal sphincter (UES). Methylene blue stained saline was not visible in any patient during pressure-controlled ventilation, but was detected in two of the 36 patients with the i-GelTM during spontaneous breathing. Conclusion: Instilling dye through the drainage tube of SGA models with a built-in drainage tube represents a useful method to examine and to compare the fluid seal of different SGA. Our protocol presented in this study proved to be an easy and reproducible approach for future studies. Furthermore, the clinical results gained during this evaluation highlight the necessity for further investigations regarding the fluid seal competencies of SGAs in humans under clinical conditions.

Share and Cite:

S. G. Russo, S. Cremer, U. Mühlhäuser, C. Eich, M. Quintel and M. Bauer, "Investigating the Fluid Seal of Supraglottic Airway Devices in Humans Using Indicator Dye via the Drainage Tube: A Potential Roadmap for Future Studies," Open Journal of Anesthesiology, Vol. 2 No. 2, 2012, pp. 18-22. doi: 10.4236/ojanes.2012.22006.

Conflicts of Interest

The authors declare no conflicts of interest.

References

[1] T. M.Cook and B. Howes, “Supraglottic Airway Devices: Recent Advances,” Continuing Education in Anaesthesia, Critical Care & Pain, Vol. 11, No. 1, 2011, pp. 1-33.
[2] T. M. Cook, G. Lee and J. P. Nolan, “The ProSealTM Laryngeal Mask Airway: A Review of the Literature,” Canadian Journal of Anaesthesia, Vol. 52, No. 7, 2005, pp. 739-760. doi:10.1007/BF03016565
[3] W. Schmidbauer, S. Bercker, T. Volk, G. Bogusch, G. Mager and T. Kerner, “Oesophageal Seal of the Novel Supralaryngeal Airway Device i-Gel in Comparison with the Laryngeal Mask Airways Classic and ProSealTM Using a Cadaver Model,” British Journal of Anaesthesia, Vol. 102, No. 1, 2009, pp. 135-139. doi:10.1093/bja/aen319
[4] S. Bercker, W. Schmidbauer, T. Volk, G. Bogusch, H. P. Bubser, M. Hensel and T. Kerner, “A Comparison of Seal in Seven Supraglottic Airway Devices Using a Cadaver Model of Elevated Esophageal Pressure,” Anesthesia & Analgesia, Vol. 106, No. 2, 2008, pp. 445-448. doi:10.1213/ane.0b013e3181602ae1
[5] J. J. Gatward, T. M. Cook, C. Seller, J. Handel, T. Simpson, V. Vanek and F. Kelly, “Evaluation of the Size 4 i-GelTM Airway in One Hundred Non-Paralysed Patients,” Anaesthesia, Vol. 63, No. 10, 2008, pp. 1124-1130. doi:10.1111/j.1365-2044.2008.05561.x
[6] A. van Zundert and J. Brimacombe, “The LMA Supreme—a Pilot Study,” Anaesthesia, Vol. 63, No. 2, 2008, pp. 209-210. doi:10.1111/j.1365-2044.2007.05421.x
[7] R. Townsend, J. Brimacombe, C. Keller, V. Wenzel and H. Herff, “Jaw Thrust as a Predictor of Insertion Conditions for the Proseal Laryngeal Mask Airway,” Middle East Journal of Anesthesiology, Vol. 20, No. 1, 2009, pp. 59-62.
[8] J. Brimacombe and C. Keller, “Prime the ProSealTM Drain Tube with Lube from a Tube!” Canadian Journal of Anaesthesia, Vol. 52, No. 3, 2005, pp. 338-339. doi:10.1007/BF03016079
[9] C. Verghese, J. Berlet, A. Kapila and R. Pollard, “Clinical Assessment of the Single Use Laryngeal Mask Airway—the LMA-Unique,” British Journal of Anaesthesia, Vol. 80, No. 5, 1998, pp. 677-679.
[10] C. Keller, J. R. Brimacombe, K. Keller and R. Morris, “Comparison of Four Methods for Assessing Airway Sealing Pressure with the Laryngeal Mask Airway in Adult Patients,” British Journal of Anaesthesia, Vol. 82, No. 2, 1999, pp. 286-287. doi:10.1093/bja/82.2.286
[11] B. Gibbison, T. M. Cook and C. Seller, “Case Series: Protection from Aspiration and Failure of Protection from Aspiration with the i-Gel Airway,” British Journal of Anaesthesia, Vol. 100, No. 3, 2008, pp. 415-417. doi:10.1093/bja/aem396
[12] N. R. Evans, S. V. Gardner and M. F. James, “ProSeal Laryngeal Mask Protects against Aspiration of Fluid in the Pharynx,” British Journal of Anaesthesia, Vol. 88, No. 4, 2002, pp. 584-587. doi:10.1093/bja/88.4.584
[13] A. Lallo, P. Drolet and M. Lacroix, “Performance of the PAXpressTM vs the ProSealTM Laryngeal Mask Airway during General Anesthesia,” Canadian Journal of Anaesthesia, Vol. 54, No. 1, 2007, pp. 28-33. doi:10.1007/BF03021896
[14] A. Timmermann, S. Cremer, C. Eich, S. Kazmaier, A. Brauer, B. M. Graf and S. G. Russo, “Prospective Clinical and Fiberoptic Evaluation of the Supreme Laryngeal Mask Airway,” Anesthesiology, Vol. 110, No. 2, 2009, pp. 262-265.

Copyright © 2024 by authors and Scientific Research Publishing Inc.