Patient dose from radiographic rejects/repeats in radiology centers of Urmia University of Medical Sciences, Iran

Abstract

Medical diagnostic X-rays are the largest manmade source of ionizing radiation received by the members of the general public. The aim of this study was to evaluate the radiographic reject/repeat rate and also to determine dose to the patients from radiographic rejects/repeats in radiology centers of Urmia University of Medical Sciences. During a 4 month period the most frequently examinations were chosen in three radiology centers. A form was designed as a reject/repeat analysis form for radiographers to complete each time a film was rejected by radiologists or repeated. The collected data were compiled at the end of each week and entered into a computer for analysis at the end of study. The results of this study showed that highest and lowest repetition rates were for pelvis, 14.01% and upper limb, 4.17%, respectively. The main reasons of repetition of radiographs were due to exposure (54%) and positioning (18%) errors. The average repeat rate in all three hospitals was 7.20%. It was found that human error has important role to repetition of radiographs. It is demonstrated that those patients having repeated radiographs received an average of 3.23 Gy·cm2. Based on the findings of this study it must be remembered that the highest repetition rate was for pelvis. Considering the radiosensitive organs related to pelvis especially in pediatric patients some special considerations must be applied for pelvis examinations.

Share and Cite:

Jabbari, N. , Zeinali, A. and Rahmatnezhad, L. (2012) Patient dose from radiographic rejects/repeats in radiology centers of Urmia University of Medical Sciences, Iran. Health, 4, 94-100. doi: 10.4236/health.2012.42015.

Conflicts of Interest

The authors declare no conflicts of interest.

References

[1] Bushberg, J.T., Seibert, J.A., Leidholdt, E.M. and Boone, J.M. (2002) The essential physics of medical imaging. Lippincott Williams and Wilkins, Philadelphia.
[2] Australian Radiation Protection and Nuclear Safety Agency (2008) Radiation protection in diagnostic and interventional radiology. Radiation Protection Series Publication No. 14.1, Victoria.
[3] Bassey, C.E., Ojo, O.O. and Akpabio, I. (1991) Repeat profile analysis in an x-ray department. Journal of Radiological Protection, 11, 179-183.
[4] Al-Malki, M.A., Abulfaraj, W.H., Bhuiyan, S.I. and Kinsara, A.A. (2003) A study on radiographic repeat rate data of several hospitals in Jeddah. Radiation Protection Dosimetry, 103, 323-330.
[5] Supe, S.J., Lyer, P.S., Sasane, J.B., Sawant, S.G. and Shirva, V.K. (1992) Estimation and significance of patient doses from diagnostic X-ray practices in India. Radiation Protection Dosimetry, 43, 209-211.
[6] Canon, C.L. (2010) McGraw-Hill specialty board review radiology. McGraw-Hill Companies, New York.
[7] International Commission on Radiological Protection (1991) Recommendations of the international commission on radiological protection. ICRP Publication 60, Oxford.
[8] Shabestani Monfared, A., Abdi, R. and Saber, M.A. (2007) Repeat analysis program in radiology departments in Mazandaran province—Iran; impact on population radiation dose. Iranian Journal of Radiation Research, 5, 37-40.
[9] Sniureviciute, M. and Adliene, D. (2005) Problems with film processing in medical X-Ray imaging in Lithuania. Radiation Protection Dosimetry, 5, 260-263. doi:10.1093/rpd/nch577
[10] Lau, S.L., Mak, A.S., Lam, W.T., Chau, C.K. and Lau, K.Y. (2004) Reject analysis: A comparison of conventional filmescreen radiography and computed radiography with PACS. Radiography, 10, 183-187. doi:10.1016/j.radi.2004.03.014
[11] Zewdeneh, D., Teferi, S. and Admassie, D. (2008) X-ray reject analysis in Tikur Anbessa and Bethzatha Hospitals. Ethiopian Journal of Health Development, 22, 63-97.
[12] International Commission on Radiological Protection (1996) Radiological protection and safety in medicine. ICRP Publication 73, Oxford.
[13] Kiljunen, T., Jarvinen, H. and Savolainen, S. (2007) Diagnostic reference levels for thorax X-ray examinations of paediatric patients. The British Journal of Radiology, 80, 452-459. doi:10.1259/bjr/60918774
[14] International Atomic Energy Agency (2007) Dosimetry in diagnostic radiology. International code of practice. IAEA Technical Reports Series No. 457, Vienna.
[15] International Commission on Radiation unites and measurements (1998) Fundamental quantities and unites for ionizing radiation. ICRU Report 60.
[16] Duggan, L., Warren-Forward, H., Smith, T. and Kron, T. (2003) Investigation of dose reduction in neonatal radiography using specially designed phantoms and LiF:Mg, Cu,P TLDs. The British Journal of Radiology, 76, 232-237. doi:10.1259/bjr/79291075
[17] United Nations Scientific Committee on the Effect of Atomic Radiation (2008) Sources and effects of ionizing radiation. UNSCEAR Report 2008.
[18] Watkinson, S., Moores, M. and Hill, S.J. (1994) Reject analysis: Its role in quality assurance. Radiography, 50, 189-194.
[19] Dunn, M.A. and Rogers, A.T. (1998) X-ray film reject analysis as a quality indicator. Radiography, 4, 29-31. doi:10.1016/S1078-8174(98)80027-8
[20] Koichi, C., Mamoru, K., Yutaka, K., et al. (2010) Radiation dose and radiation protection for patients and physicians during interventional procedure. Journal of Radiation Research, 51, 97-105. doi:10.1269/jrr.09112
[21] International Atomic Energy Agency (1996) International basic safety standards for protection against ionizing radiation and for the safety of radiation source. IAEA Safety Series No. 115, Vienna.
[22] Nixon, P.P., Thorogood, J., Holloway, J. and Smith, N.J. (1995) An audit of film reject and repeat rates in a department of dental radiology. The British Journal of Radiology, 68, 1304-1307. doi:10.1259/0007-1285-68-816-1304
[23] Peer, S., Peer, R., Walcher, M., Pohl, M. and Jaschke, W. (1999) Comparative reject analysis in conventional film screen and digital storage phosphor radiography. European Radiology, 9, 1693-1696.
[24] Chu, W.K., Ferguson, S., Wunder, B., Smith, R. and Vanhoutte, J.J. (1982) A Two-year reject/retake profile analysis in paediatric radiology. Health Physics, 42, 53-59. doi:10.1097/00004032-198201000-00005
[25] Adler, A., Carlton, R. and Wold, B. (1992) An analysis of radiographic repeat and reject rates. Radiologic Technology, 63, 308-314.
[26] Mustafa, A.A., Vasisht, C.M. and Sumarasekara, S.J. (1987) Analysis of wasted x-ray films: Experience in two Kuwait hospitals. The British Journal of Radiology, 60, 513-515. doi:10.1259/0007-1285-60-713-513
[27] Hart, D., Hillier, M.C. and Wall, B.F. (2000) Doses to patients from medical X-ray examinations in UK. NRPB-W14, Chilton.
[28] Müller, M., Heicappell, R., Steiner, U., Merkle, E., Aschoffm, A.J. and Miller, K. (1998) The average dose-area product at intravenous urography in 205 adults. The British Journal of Radiology, 71, 210-212.
[29] Clark, P.A. and Hogg, P. (2003) Reject/repeat analysis and the effect prior film viewing has on a department’s reject/repeat rate. Radiography, 9, 127-137. doi:10.1016/S1078-8174(03)00036-1
[30] Committee on the Biological Effects of Ionizing Radiation (1990) Health effects of exposure to low levels of ionizing radiation. BEIR V, Washington DC.
[31] McEntee, M.F. and Kinsella, C. (2010) An examination of practice during radiography of the clavicle. Radiography, 16, 125-130. doi:10.1016/j.radi.2009.12.002

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.