Assessment of Modeling Collimator Designs for Gamma-Ray Transmission of Uranium Oxide Spectrometry Using HPGe Detectors ()
1. Introduction
In several scientific disciplines, gamma-ray spectroscopy is a commonly utilized technique. It is utilized in a variety of fields, including nuclear technology and fundamental physics research. The application of gamma-ray spectroscopy in non-destructive nuclear fuel inspections is used e.g. gamma scanning and Gamma Emission Tomography [1] . In particular, the latter is increasingly considered for use in nuclear fuel tests as well as in nuclear safeguards [2] - [7] . In these situations, the measuring devices consist of a high energy resolution detector and a collimator, where the collimator enables the investigation of well-defined fuel zones. The geometry and material composition of the collimator should be carefully considered during the beginning stages of creating a spectroscopic system since they have a significant impact on the system’s performance. MC simulations are frequently used to produce outcomes with the potential for high accuracy [8] [9] [10] . For instance, for an isotopically emitting point source positioned directly in front of the collimator opening, the likelihood of a gamma ray reaching the opposite opening is on the order of 10−9 for a collimator with a square slit that is 80 cm long and 0.1 cm wide. The probability of penetration can be substantially smaller for the parts of a fuel object that are not directly in front of the opening or further away. As a result, MC approaches need a lot of photons to be sampled, which can be time-consuming if no special variance reduction techniques are used, such as those proposed by. The Monte Carlo radiation transport code MCNP6 is used to model a mono-directional source, and correction factors are added to account for the effects of a cylindrical aperture and solid-angle effects of isotropic emission [10] [11] .
The nondestructive assay (NDA) of radioactive waste detection uses a technique called Tomographic Gamma Scanning (TGS), which is relatively recent [1] [2] [3] [4] [5] . It helps to resolve the issue of invalid attenuation correction brought on by the unequal distribution of the sample medium in industrial CT imaging technologies. As a result, it increases the precision of the information contained in the non-uniform analysis of radioactive materials in the measurements of gamma ray spectroscopy [6] [7] [8] [9] . With a high sample throughput and sensitivity, the TGS approach seeks to produce precise measurements of radionuclides with low specific activity. The image quality, in the sense it is generally understood, is of little concern beyond its effect on assay accuracy. In some research outcomes, a TGS design study was conducted to enhance the functionality and other tomographic assay systems. The study was based on computer simulations. The investigation of the distribution of the various gamma-ray emitting isotopes typically present in irradiated nuclear fuel is made possible by the use of a high-resolution HPGe detector in the measurements, which enables the collection of detailed gamma-ray spectra from which any peak present may be chosen for tomographic reconstruction. The number of counts gathered in a specific gamma-ray peak is significantly more than the number of counts contained in the peak in a single spectrum because gamma spectra are recorded at between 8000 and 25,000 points surrounding the fuel. As a result, usable images can be created from even peaks in a single spectrum that has a relatively small number of counts [12] [13] .
In this article, we describe the findings of a collimator design study that was conducted to enhance the functionality of safeguards measurements and tomographic assay systems. The study was based on Monte Carlo simulations to calculate the HPGe detector efficiency. Usually, the absolute efficiency is determined by the pulse-height tally (F8) of each photon emitted from the source. The focus of this work is on the associated problems with collimator design.
2. Theoretical Model
The count rate, CR, in the detector of a spectroscopic system can be described as a function of the source activity A, the emission intensity, and the gamma-ray energy per decay of the gamma line
, the geometric efficiency
, and the detector intrinsic full-energy peak efficiency,
, as expressed in Equation (1)
(1)
By using this expression, the amount of counts in the gamma-ray spectrum’s full-energy peak can be used to experimentally evaluate a sample’s activity. Notably, depending on the measurement settings, a number of correction factors, such as self-attenuation in the object, count rate-related effects, or real coincidence summing, may be necessary for accurate results. The geometric efficiency
, meaning the probability per photon to reach the detector, is in this case determined by the collimator slit dimensions and the collimator-source distance.
Typically, just a small portion of the source plane is covered by the collimator slit. Because of this, it is more useful to express the count rate intensity based on the source’s planar activity concentration, adapted from [14] , as shown in Equation (2) below:
, (2)
We have different geometrical designs for testing which are the (cylinder & rectangle) shapes with different dimensions that are carried out. The cylindrical shapes are designed with different diameters and heights while the rectangular shape is designed with different lengths, widths, and heights. We assumed six designs as shown in Table 1. The first three designs with design ID: Dx, Dy, and Dz refer to cylindrical shapes with different diameters and heights while, the other three designs that are with design ID: Dx', Dy', and Dz' refer to the rectangular shape with different dimensions in height, length, and width. Each design in the cylindrical shape with a fixed diameter is simulated with different four heights (H1, H2, H3, and H4).
3. Simulation Setups
The MCNP code can only analyze an issue in the form that it is given; it cannot tell when the wrong material has been specified or when the geometry has been inaccurately modeled. Keep in mind that without knowledge of the problem’s context, the standard of the solution, and a reasonable expectation of the outcome and its associated confidence interval with MCNP or any other code is useless [15] [16] [17] .
In this section, the simulated design for the collimator is described with detailed dimensions and configuration. Figure 1 describes the simulated setup where the standard nuclear material (SNM), collimator, and HPGe detector are
Table 1. Detailed dimensions of collimator designs.
Figure 1. Collimator design (cylindrical shape).
on the same alignment and the distance between each one is 3 cm. The measuring system is composed of: A high purity planar HPGe detector (Canberra; model GL0515R), active area = 500 mm2) 540 eV FWHM resolution at 2 keV, a Spectroscopy Pre-amplifier [model 2002 CPSL], a Cryostat (model 7905 SL-5) liquid nitrogen Dewar, an adjustable High-Voltage Power Supply (HV-PS) Inspector Multi-Channel Pulse-height Analyzer (MCA) module (model 1200 IMCA) and built-in amplifier, a 100-MHZ Wilkinson ramp Amplitude-to-digital Converter (ADC), and an online microcomputer data acquisition module for the control and analysis [18] [19] . The SNM compose of a compact powder of Uranium oxide (U3O8) with a total weight of 200.1 gm and covered with Al- can [20] . The experimental work is applied by using the SNM sample with an enrichment 4.46% and the observed energy line is at 185.7 keV. The whole shape of the can is simulated in the input file of the MCNP code. Figure 1, Figure 2 presents the whole configuration of the simulated set up including the different shapes of collimators (cylindrical and rectangle) with different dimensions in diameter (D) and length (L), for cylindrical shape while the rectangular shape with different length, width (W) and height (H).
Figure 2. Collimator design (rectangular shape).
Where, the above parameters are required for the simulation and the detailed data are illustrated in Table 2. The collimator designs that take design ID: Dx1, Dx2 and Dx3 have the same diameter which is equal to 1 cm with various height (H) that are taking the values from 5 to 20 cm. Similarly, the designs with design ID: Dy1, Dy2 and Dy3 that have the same diameter 2 cm with different heights and finally, the collimator designs Dz1, Dz2 and Dz3 have the same diameter 3 cm. on the other side, the designs that take rectangular shapes such as; Dx1', Dx2' and Dx3' have the same length (L) = 1 cm and width (W)= 1 cm with different H from 5 to 20 cm. Also, Dy1, Dy2, and Dy3 have L × W = 1.5 × 1.5 with a height from 5 - 20 cm.
4. Results
The simulation of the configuration shows that varying dimensions for collimator designs. The magnitude of the gamma current that reaches the detector exceeds several orders of magnitude. A good overall agreement was noted with the MCNP6simulations, including a perfectly performing collimator, and taking into account significant parameters such as; solid angle, attenuation, and geometrical efficiency that depends on the distance between the source and collimator and the dimensions of the collimator itself. Note that the simulations have been run with the same starting seed, to better observe the trend of the intensity variations due to the geometry of the setup. The input files are created for each collimator design to estimate pulse height distribution (F8). Figure 3 shows the effect of each design on the pulse height distribution where Dz denotes to change of height values for the cylindrical design with a fixed diameter 1 cm and Dy describes the change of height values for the cylindrical design with a fixed diameter 2 cm and Dz describes the change of height values for the cylindrical design with a fixed diameter 3 cm. Also, the identical idea is performed for the rectangular design Dx', Dy', Dz'.
Table 2. The required parameters that needed in MCNP6 input files.
Figure 3. Correlation between collimator height with F8 tally.
The results of the MCNP6 validation simulations including the penetrating gamma component crossing the collimator bulk material, show that different values for F8, are systematically underestimating the pulse height distribution observed by the detector. This might have been expected since the variance reduction techniques have been used to overcome the ling estimation time in the input file and give us more precise results that should be obtained. In Figure 4, the Difference between the experimental and simulated results in both shapes is presented. We found here that largest differences are in rectangular collimator shapes that is because of the corners of the collimator which cause multiple scattering that leads to decreasing in interested photon energy.
Figure 4. Difference results between the experimental and simulation.
5. Conclusion
Monte Carlo simulations have been carried out to design a collimator to improve the performance of the measurement system. The simulation results reveal that the cylindrical shape design for the collimator with ID: Dy has a diameter of 2 cm and a height range from 5 to 20 cm and it gives the acceptable difference in the range of 3.5% - 5% in the optimal shape. The optimal shape of the detector collimator was determined with the same collimator aperture radius and height. The rectangular shape of the collimator is considered the worst that’s because of the corners of the collimator which cause multiple scattering that leads to a decrease in interested photon energy. Since an HPGe detector was used in the measurements the collected gamma-ray spectra had high resolution and contained many peaks which are therefore available for tomographic reconstruction and analysis. The improvement of spatial resolution for the spectrum utilising the chosen collimator designs is the suggested direction for future research.