The Effect of the Trace Elements Concentrations on the Cancerous and Healthy Tissues in Radiotherapy ()
1. Introduction
Cancer is considered as one of the most deadly diseases and the leading causes of death worldwide [1]. Cancer treatment is performed by various methods such as surgery, chemotherapy, brachytherapy and radiotherapy or the application of these methods together. Approximately 50% of all cancer patients undergo external radiotherapy (ER) using photons [2] [3] [4]. Linear accelerators used in external beam radiation therapy treatments enable patients to be irradiated at different dose rates. The main aim of radiotherapy is to give the highest dose to the tumor and to give the lowest dose to the healthy tissue around the tumor. It is important to confirm the accuracy of the dose given to the patient during treatment planning by using the dose distribution algorithms. The particle transport calculations in the presence of an internal or external source, and the energy stored in the tissue can be determined by Monte Carlo (MC) calculations. In many studies, MC simulation packages have been used for radiation dosimeter calculations; such as OMEGA [5], MCNP4C [6] [7], EGS [8] [9]. In this research, the experimental results were simulated with the open-source code called Geant4 Application for Emission Tomography (GATE) MC simulation based on the GEANT4 toolkit (http://www.opengatecollaboration.org/). Geant4 is a software toolkit for the simulation of the interaction of particles with matter, and its application areas include high-energy physics experiments, astrophysics and astroparticle physics, nuclear physics, space science, medical physics and medical imaging, radiation protection [10] [11]. GATE plays a key role in the simulation of the radiotherapy experiments, PET, SPECT studies and design of new medical imaging devices [12] [13]. Recent studies have emphasized the importance of trace elements in the investigation about the possible causes of cancer [14] [15] [16] [17]. Trace elements have a significant effect as a component of many enzymes in all biological systems [18]. Although trace elements constitute a minor part of living tissues, they are important for vital processes. Trace element levels were determined to have deficiency or excess concerning the normal values in some diseases including cancer [19] [20]. Trace elements have different concentrations in healthy and cancerous tissues due to biological changes induced by the disease. The concentrations of trace elements in healthy and cancerous tissues or fluids have been recently obtained by using many experimental techniques such as particle-induced X-ray emission (PIXE), X-ray fluorescence analysis (XRF) and its total reflection geometry method (TRXRF), atomic absorption spectrometry (AAS) and neutron activation analysis (NAA) [21] [22] [23] [24] [25].
The concentration of trace elements has a noticeable effect on dose distributions at brachytherapy treatment (an internal radiation) which is used the low energy photon sources [26] [27]. In brachytherapy, the dosimetric impact of trace elements has been studied for normal and cancerous tissues using low energy photon sources with Monte Carlo (MC) calculations [28]. The results showed that in the presence of trace elements, dose distributions varied depending on the atomic number and fraction of the elements in tissue.
The effect of the tissue composition on dose distribution was also investigated by using electron beams in radiotherapy [29]. Ghorbani et al. [29] showed that differences in dose distribution were not significant in various soft tissues and tissue-equivalent materials. However, due to the differences in the composition of the materials, it has been proposed to be investigated the uncertainties in the calculations.
This study aimed to determine the effect of the trace elements concentrations on dose distribution in healthy and cancerous prostatic tissues in radiotherapy. The trace elements concentrations given in the literature were mixed homogeneously inside the water phantom. Here, we used the trace element compounds completely soluble in the water. The experiment was performed for 6 MV photon beams of the Elekta Synergy Platform Linear Accelerator. The percent depth-dose distributions (PDD) of the concentration of each element were measured, and the results were compared with the simulations using GATE/GEANT4 code.
2. Materials and Methods
2.1. Percentage Depth Dose Measurements
The experiments were performed by Elekta Synergy Platform Linear Accelerator in the University of Health Sciences, Dışkapı Yıldırım Beyazıt Training and Research Hospital, Department of Radiation Oncology. This linear accelerator delivers two different photon energies at 6 MV and 18 MV, and five electron energies at 6 MeV, 9 MeV, 12 MeV, 15 MeV, and 18 MeV. Firstly, the profile analysis (symmetry, flatness, penumbra, etc.) was performed using IBA Blue Phantom2 according to standard protocols. The technical specifications of IBA Blue Phantom2 are given in Table 1. Measurements were carried out with an electrometer, CC13 and FC65P ion chambers (IBA Dosimetry, Nuremberg, Germany). Ion chambers were calibrated by Turkey Atomic Energy Agency Secondary Standard Dosimetry Laboratory (ISDL).
![]()
Table 1. Technical specifications of IBA Blue Phantom2.
*PMMA: Polymethyl methacrylate, **L: Length, W: Width, H: Height.
Water tank made from PMMA similar to Blue Phantom2 was used for each measurement. The water phantom dimension was 30 × 30 × 22 cm3 with 5 mm thickness and 20 liters volume. The outer surface of the water phantom was marked up to 20 cm with 1 cm intervals. 200 MU was delivered at 6 MV photons. The source to surface distance was 100 cm and the irradiation field sizes were 10 × 10 cm2 and 20 × 20 cm2. FC65P ion chamber was placed in a solid water phantom (RW3, IBA; Schwarzenbruck, Germany). Then, the water tank was located at the top of the solid water phantoms. Pure water and all chemical elements used in the measurements were obtained from Ankara University Chemistry Department Central Warehouse.
Firstly, the tissue phantom ratio for depths of 20 and 10 cm (TPR20/10) was obtained for 10 × 10 cm2 field size to validate the simulation for 6 and 18 MV photon energies [30]. The beam quality is specified by TPR20/10 for medical linear accelerators in high energy photons. The TPR20/10 can be delivered from the following equation:
PDD20/10 is the ratio of percent depth dose at 20 and 10 cm for a field size of 10 × 10 cm2 with 100 cm SSD. Then, the simulated PDD for each trace element concentration were compared with the experimental results for 10 × 10 cm2 and 20 × 20 cm2 field sizes in 6 MV photon energies. The experimental setup used in the study is shown in Figure 1.
The FC65P ion chamber and some of the elements used in this study and the experimental setup for water and iron are shown in Figure 2. After reviewing the literature on trace element concentrations, the trace elements having the highest concentration in cancerous prostate tissues were determined
![]()
Figure 1. The schematic representation of linear accelerator and water tank. The components are not to scale.
[23] [31] [32] [33] [34]. The PDD measurements were at first performed with pure water, which is a tissue-equivalent material. The concentrations for each trace element were calculated according to water phantom volume. Water-soluble compounds of these elements were given in Table 2 (https://periodic-table-of-elements.org/SOLUBILITY, Accessed: 10.05.2020). The trace element concentrations of healthy and cancerous prostate tissues given in the literature were used to obtain a mixture of water [35]. Then, the PDD measurements were carried out for each trace element compound.
2.2. GATE Simulations
For dosimetry related applications in radiation therapy, the GATE v8.1 release was used to simulate of Elekta Synergy Platform Linear Accelerator at 6 MV photon energies. The water phantom material and size the same as the experiment was modelled in the simulation which is called a mainbox. The PDD curves were obtained for a 10 × 10 cm2 and 20 × 20 cm2 fields at 100 cm SSD by using dose of 20 mm × 20 mm × 10 mm and Dose Actor at these energies. The Dose Actor scores an energy deposited, distribution of dose and the associated
![]()
Table 2. The trace element mass fraction (mg/kg, dry mass basis) in the healthy and cancerous prostate tissues used in this study and their compounds which are completely soluble in water.
*Zaichick and Zaichick [35].
statistical uncertainty in any volume [10]. We determined the Standard physics list with option 3 for photons, e− and e+ [36]. Four Kill Actors were defined at the surface of the phantom. The Cut in Region was set to 1 mm in the world and to 0.1 mm in the phantom for electrons, positrons and photons. The chemical compound of each trace element and their concentration in water were classified inside the Gate material list. The number of histories for all simulation was 3 × 109.
3. Results and Discussion
First, the flatness and symmetry of the instrument were determined by using the IBA Blue Phantom2 water phantom. In Linac, the flatness value should be less than 3% and the symmetry value should be less than 2%. These values can be obtained in the largest field size at 100 cm SSD and 10 cm depth [37]. TPR20/10 was calculated for 10 × 10 cm2 field size at 6/18 MV energies photons and compared with simulation results. The differences between simulation and measurement for 6 MV and 18 MV photons are 1.75% and 1.82% respectively, as shown by the values in Table 3. Data obtained with MC simulations presented less than 3% uncertainty. Similarly, data from all experimental measurements presented less than 3% uncertainty.
The difference between the experimental and simulation was calculated by
where
and
are doses at the position r of the experimental and simulated curves and
is the maximum dose of the experimental curve. SPSS software (Version 22.0. SPSS. Inc., USA) was used to determine the statistical significance of the difference between experimental and simulation data by means of paired t test. The p value of less than 0.05 was considered statistically significant. Dose differences with p values are shown in Table 4.
Teixeira et al. [38] created the phase space of the Novalis Classic linear accelerator at 6 MV energy performed the GATE simulation program and compared PDD and dose profiles with experimental data for 10 × 10 cm2 and 3 × 3 cm2 radiation field. They found that the TPR20/10 difference between simulation and measurement was about 1.5% for 6 MV. In our study, the difference between simulation and experiment was calculated at 1.75%. In Figure 3 and Figure 4 the experimental measurements performed for pure water were compared with the simulated results at 6/18 MV photons, at 10 × 10 cm2 and 20 × 20 cm2 field sizes. The PDD distributions for the concentrations of trace elements in healthy and cancerous prostate tissues were measured by using the trace element mass fraction as indicated in Table 2. The same conditions were defined and then simulated in GATE. ICRU 24 [39] recommends that the uncertainty in the dose given in radiotherapy should not exceed ± 5%. In this study, the difference between measured and simulated results for the pure water measurement did not exceed ±2%.
Grevillot et al. [40] measured PDD and dose profiles at 6 MV photons by using Elekta Precise Linac device and simulated with GATE program. They also found the dose differences between simulation and measurements approximately
![]()
Table 3. TPR, flatness and symmetry values for 6 and 18 MV photons.
![]()
Table 4. Dose differences between pure water, CTCx4 and simulation.
![]()
Figure 3. The comparison of experimental and simulated PDD’s values for pure water at 10 × 10 cm2, 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 4. The comparison of experimental data and simulation results for pure water at 10 × 10 and 20 × 20 cm2 field sizes for 18 MV photons.
1% - 2%. In similar studies, PDD and dose profiles measurements were performed for the various Linac devices and compared with different simulation codes. The results were consistent with each other [40] [41] [42] [43] [44]. The concentrations of the cancerous tissue (CTC) were increased the four times to investigate the effect of the maximum concentration on the dose distribution. In Figures 5-13, the simulated results of CTCx4 were compared with each experimental data set for healthy tissue (HTC) and cancerous prostate tissue concentrations. In the experimental results for each element concentration, there was no difference between the PDDs of the healthy tissue and cancerous tissue. Although the cancerous tissue concentration was increased by four times, it was found that the distributions were no different from the dose distributions obtained for cancerous, even healthy tissue concentrations. High energy photons primarily interact through Compton scattering, which is Z-independent. Therefore, the change in trace element concentrations in healthy and cancerous tissue did not affect the dose distribution. Since there were many experimental and simulated results, only the simulated data for CTCx4 were given in figures.
Ghorbani et al. [29] studied the effect of the tissue composition on dose distribution for 8/12/14 MeV electron beams in radiotherapy. The various soft tissues and tissue-equivalent materials were simulated using MCNPX MC code for a Siemens Primus linear accelerator. There were no differences in dose distributions in various soft tissues and tissue-equivalent materials. Ghorbani et al. [45]
![]()
Figure 5. Comparison of experimental data and simulation results for pure water and AlCl3∙6H2O at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 6. Comparison of experimental data and simulation results for pure water and KBr at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 7. Comparison of experimental data and simulation results for pure water and CaCl2∙2H2O at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 8. Comparison of experimental data and simulation results for pure water and FeCl3∙6H2O at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 9. Comparison of experimental data and simulated results for pure water and KCl at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 10. Comparison of experimental data and simulation results for pure water and MgCl2 at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 11. Comparison of experimental data and simulation results for pure water and MnCl2∙4H2O at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 12. Comparison of experimental data and simulation results for pure water and NaCl at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 13. Comparison of experimental data and simulation results for pure water and ZnSO4∙7H2O at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
![]()
Figure 14. Comparison of the experimental data and simulation results, for pure water and mixture of nine elements at 10 × 10 and 20 × 20 cm2 field sizes for 6 MV photons.
also searched the effect of the soft-tissue composition on dose distribution using Siemens Primus linear accelerator at 6 MV photons. The soft tissue and three types of tissue-equivalent materials were also simulated using MCNPX MC code for Siemens Primus linear accelerator. They found minor differences between dose distributions in various soft tissues and tissue-equivalent materials.
White et al. [28] determined the dose distribution for the trace element concentrations at healthy or cancerous human tissues with low energy photon sources in brachytherapy. They simulated the dose distribution with Geant4 v9.3 and found that the different trace element concentration between healthy and cancerous prostate tissues affected the dose distribution and it should not be ignored.
In Figure 14, the experimental data performed with pure water and mixture of nine elements (CTCx4) for 6 MV photons, at 10 × 10 cm2 and 20 × 20 cm2 field sizes were compared with the simulation.
4. Conclusion
To investigate the effect of trace element concentrations in tissue on dose distribution, the experimental and simulated PDD values for pure water, healthy and cancerous prostate tissues were obtained at 6 MV photon energy. The experimental values of PDD were in a good agreement with the simulated data using GATE simulation code. There is a difference of less than 2% between the measured and simulated results. The experimental results and simulations were presented an uncertainty lower than 3%. As expected, it is seen that the difference between the trace element concentrations of healthy and cancerous tissues did not affect the dose distribution at high-energy photons. This is expected and well known result. We believe that this in vitro study is important for proving the reliability of the dose given in radiotherapy treatment once again.