Attention Detection System Based on the Variability of Heart Rate

HTML  XML Download Download as PDF (Size: 1781KB)  PP. 54-70  
DOI: 10.4236/jst.2019.94006    1,027 Downloads   2,965 Views  Citations

ABSTRACT

Attention is a cognitive and psychobiological variable that lacks objective measures in natural contexts of formal learning. However, it can be objectively measured from the heart autonomic activity. In the present study, the validity and reliability of a system for recording heart rate variability are demonstrated. Twelve prototypes were created and paired with standard electrocardiogram. Long segments with cardiac pulse recordings were used within ten minutes. A highly significant correlation index (p < 0.01) and a significantly high Cronbach’s alpha were obtained. In the variance analysis, all data analyzed presented a value of p < 0.05. It is concluded that there is no significant variation between paired systems; therefore, the attention detection system based on heart rate variability is valid and reliable.

Share and Cite:

Avila, U. , Braga, I. , Campos, F. and Nafital, A. (2019) Attention Detection System Based on the Variability of Heart Rate. Journal of Sensor Technology, 9, 54-70. doi: 10.4236/jst.2019.94006.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.