Periodic Measurements of Passive Proton Beam Width Using Radiochromic Film in Fixed Gantry System

HTML  XML Download Download as PDF (Size: 2957KB)  PP. 193-203  
DOI: 10.4236/ijmpcero.2019.84017    466 Downloads   1,068 Views  

ABSTRACT

Background and Aim: Irradiation methods such as double scattering method and spot scanning method have been used in proton beam treatment devices. In the scattering method, a ridge filter or a range modulation wheel is used to create a spread-out Bragg peak, but the distribution at the patient position may change due to positional deviation of the incident beam. Therefore, assessment of the incident position of the beam is very important even in the scattering method. To investigate the width and distribution of the proton beam before entering the RMW, a radiochromic film was installed at the outlet of the transport pipe and the entrance of the profile-monitoring detector. Methods: In this study, the distributions of the beam at the exit of the transport pipe and the entrance of the monitor detector were measured using films. The beam width was measured from the full width at half maximum of the profile obtained from the distribution. Measurements were conducted every month for 10 months. Results: Beams of widths ranging from 1.82 to 2.30 mm in the horizontal direction and 4.25 to 5.33 mm in the vertical direction were outputted from the exit of the transport pipe. Beams of widths ranging from 2.16 to 2.67 mm in the horizontal direction and 4.06 to 5.31 mm in the vertical direction were outputted from the entrance of the monitor detector. The maximum width fluctuation for 10 months was 0.55 mm in the horizontal direction and 1.26 mm in the vertical direction at the entrance of the monitor detector. Conclusions: The distribution was obtained before the proton beam was scattered by the scatterer, and then we propose a method to periodically measure and monitor the changes in the beam distributions every month.

Share and Cite:

Shimozato, T. , Yasui, K. , Kinou, H. and Komatsu, F. (2019) Periodic Measurements of Passive Proton Beam Width Using Radiochromic Film in Fixed Gantry System. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 8, 193-203. doi: 10.4236/ijmpcero.2019.84017.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.