Preliminary Monte Carlo Investigation of Using Ir-192 as the Source for Real Time Imaging Purpose

HTML  XML Download Download as PDF (Size: 2565KB)  PP. 21-30  
DOI: 10.4236/ijmpcero.2017.61003    1,674 Downloads   2,651 Views  Citations
Author(s)

ABSTRACT

The purpose of this study is to investigate the potential use of Ir-192 as the source for real time imaging during HDR (High Dose Rate) brachytherapy treatment. Phantom measurement was performed to determine outside of the body dose. Monte Carlo code, EGSnrcMP egs_inprz, was used for the simulation to calculate the outside of the body x-ray signal for CT reconstruction. Matlab code was developed to reconstruct the Ir-192 source and for 3D visualization in order to assess reconstructed CT resolution, signal-to-noise ratio, and imaging dose information. The measured dose was 0.67 ± 0.04 cGy, which was comparable to the Monte Carlo simulation result 0.71 ± 0.20 cGy. The reconstructed source diameter dimension was 1.3 mm compared with 1.1 mm for the real source dimension. The signal-to-noise ratio was 19.91 db following de-noising. Source position was within a 1 mm difference between programmed and simulated results. Although the Ir-192 signal is weak for CT imaging, it is possible to use it as a CT imaging x-ray source for HDR treatment localization, verification and dosimetry purposes. Further study is needed for the detailed design of an outside of the body CT-like device for use in brachytherapy imaging.

Share and Cite:

Shi, C. and Wang, B. (2017) Preliminary Monte Carlo Investigation of Using Ir-192 as the Source for Real Time Imaging Purpose. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 6, 21-30. doi: 10.4236/ijmpcero.2017.61003.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.