Limitations of On-Wafer Calibration and De-Embedding Methods in the Sub-THz Range

HTML  Download Download as PDF (Size: 229KB)  PP. 25-29  
DOI: 10.4236/jcc.2013.16005    5,481 Downloads   7,930 Views  Citations

ABSTRACT

This paper investigates frequency limitations of calibration and de-embedding techniques for S parameter measurements. First, the TRL calibration methods are analysed and the error due to the probe movement when measuring the different line lengths is quantified, next the coupling between the probe-heads and the wafer surface is investigated and finally an upper frequency validity limit for the standard Open-Short de-embedding method is given. The measured results have been confirmed thanks to the use of an electro-magnetic simulator.

Share and Cite:

Potereau, M. , Raya, C. , Matos, M. , Fregonese, S. , Curutchet, A. , Zhang, M. , Ardouin, B. and Zimmer, T. (2013) Limitations of On-Wafer Calibration and De-Embedding Methods in the Sub-THz Range. Journal of Computer and Communications, 1, 25-29. doi: 10.4236/jcc.2013.16005.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License