Developed New Procedure for Low Concentrations of Hydrazine Determination by Spectrophotometry: Hydrazine-Potassium Permanganate System

HTML  Download Download as PDF (Size: 408KB)  PP. 98-102  
DOI: 10.4236/jasmi.2012.22018    7,402 Downloads   16,748 Views  Citations

ABSTRACT

An indirect, sensitive and accurate method for the determination of trace amounts of hydrazine is described. In this proposed the spectrophotometric method is based on its reduction properties of hydrazine with a known concentration of potassium permanganate to reduce the colour. The absorbance of unreduced permanganate is measured the colour difference at different wavelengths 546 and 526 nm which show an absorption spectrum with hydrazine. Hydrazine can be determined in the range of 100 - 700 μg/ml with correlation coefficient of 0.999 and relative standard deviation 1%. The method is successfully applied for the determination of hydrazine in water streams in nuclear reactors/purex process/boiler water and polluted water samples.

Share and Cite:

S. Ganesh, F. Khan, M. Ahmed, P. Velavendan, N. Pandey and U. Mudali, "Developed New Procedure for Low Concentrations of Hydrazine Determination by Spectrophotometry: Hydrazine-Potassium Permanganate System," Journal of Analytical Sciences, Methods and Instrumentation, Vol. 2 No. 2, 2012, pp. 98-102. doi: 10.4236/jasmi.2012.22018.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.