Simple Method for Preparing Glucose Biosensor Based on Glucose Oxidase in Nanocomposite Material of Single-Wall Carbon Nanotubes/Ionic Liquid

HTML  Download Download as PDF (Size: 532KB)  PP. 54-59  
DOI: 10.4236/jasmi.2012.22011    5,627 Downloads   11,797 Views  Citations

ABSTRACT

Based on electric conductivity and wide potential window of ionic liquid (IL) and electric property of single-wall car- bon nanotubes (SWCNTs), composite material of IL-SWCNTs was prepared, glucose sensor was built with this mate-rial for immobilizing glucose oxidase (GOx). It showed good response, sensitivity and stability for long time for glu-cose detection. Linear range for the detection of glucose was from 0.5 × 10–6 M to 12 × 10–6 M while detection limit was 6.26 × 10–8 M (S/N = 3).

Share and Cite:

W. Wang, G. Yin, X. Ma and J. Wan, "Simple Method for Preparing Glucose Biosensor Based on Glucose Oxidase in Nanocomposite Material of Single-Wall Carbon Nanotubes/Ionic Liquid," Journal of Analytical Sciences, Methods and Instrumentation, Vol. 2 No. 2, 2012, pp. 54-59. doi: 10.4236/jasmi.2012.22011.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.