A method for retrieving the waveform of the pressure pulsations from the output of an electronic oscillometer

Abstract

In the most common version of an oscillometric blood pressure monitor, the output from the pressure transducer, Y(t), is split into two parts, and used for separate determinations of the pressure inside the pneumatic cuff and its fluc-tuating part; the latter is derived by sending Y(t) to a high-pass filter (HPF) and amplifying the fil-tered part to obtain the oscillometric signal O(t). Using a typical HPF-amplifier combination, we show that if p(t), the pulsatile part of the cuff pressure, is defined to be a train of positive-going pulses, O(t) turns out to be rather close but not identical to dp/dt, and to demonstrate that one can easily retrieve p(t) from a record of O(t). This means that, with a small modification, the instrument can provide both p(t) and dp/dt; the practical advantages of this demonstration are pointed out.

Share and Cite:

Naqvi, K. , Jávorfi, T. and Vellani, C. (2008) A method for retrieving the waveform of the pressure pulsations from the output of an electronic oscillometer. Journal of Biomedical Science and Engineering, 1, 75-78. doi: 10.4236/jbise.2008.12012.

Conflicts of Interest

The authors declare no conflicts of interest.

References

[1] Hill, L. (1898) On rest, sleep, and work and the concomitant changes in the circulation of the blood. Lancet, 151, 282–285.
[2] Erlanger, J. (1916) Studies in blood pressure estimations by indirect methods: I. The mechanism of the oscillatory criteria. Am. J. Physiol., 39, 401–446.
[3] Burch, G.E. and DePasquale, N.P. (1962) Primer of Clinical Measurement of Blood Pressure, C. V. Mosby Ed. St. Louis CO 1962, 15儃36.
[4] Corall, I.M. and Strunin, L. (1975) Assessment of the Von Recklinghausen oscillotonometer. Anaesthesia, 30, 59–66.
[5] Mauck, G.W., Smith, C.R., Geddes, L.A. and Bourland, J.D. (1980) Meaning of the point of maximum oscillations in cuff pressure in the indirect measurement of blood-pressure刼Part II. Trans ASME, 102, 2833.
[6] Geddes, L.A., Voelz, M., Combs, C., Reiner D. and Babbs, C.F. (1982) Characterization of the oscillometric method for measuring indirect blood-pressure. Ann. Biomed. Eng., 10, 271–280.
[7] Chua, C.S. and Hin, S.M. (2005) Digital blood pressure meter. Freescale Semiconductor, (http://www.freescale.com/files/sensors/ doc/app_note/AN1571.pdf.)
[8] Middelhoek, S. and Noorlag, D.J.W. (1981) Silicon micro-transducers. J. Phys. E: Sci. Instrum., 14, 1343–1352.
[9] Silas, J.H., Barker, A.T. and Ramsay, L.E. (1980) Clinical-evaluation of dinamap-845 automated blood-pressure recorder. Br. Heart J., 43, 202–205.
[10] Amoore, J.N. (2006) Extracting oscillometric pulses from the cuff pressure: does it affect the pressure determined by oscillometric blood pressure monitors? Blood Press. Monit., 11, 269–279.
[11] Baker, P.D., Orr, J.A., Westenskow, D.R. and Egbert, T.P. Method for determining blood pressure utilizing a neural network. U.S. Patent 5,339,818, August 23, 1994.
[12] Jazbinsek, V., Luznik, J. and Trontelj, Z. (2005) Non-invasive blood pressure measurements: separation of the arterial pressure oscillometric waveform from the deflation using digital filtering. IFBME proceedings of EMBEC’05, (http://fizika.imfm.si/ jazbinsek/konf2005/2268F.pdf)
[13] Naqvi, K.R., Parigi, L., Vellani, C.W. and Kumar, S. (2008) Noninvasive method for determining blood pressure and contours of arterial and volume pulses. JBiSE,1,79-84.
[14] Brinton, T.J., Cotter, B., Kailasam, M.T., Brown, D.L., Chio, S., O’Connor, D.T. and DeMaria, A.N. (1997) Development and validation of a noninvasive method to determine arterial pressure and vascular compliance. Am. J. Cardiol., 80, 323–330.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.