Generation of Arbitrary Pressure Pulsation of Wide Frequency Range for Flow Meter Testing in a Laminar Gas Pipeline

Abstract Full-Text HTML Download Download as PDF (Size:2801KB) PP. 125-137
DOI: 10.4236/jfcmv.2014.24015    2,609 Downloads   3,120 Views  

ABSTRACT

The present paper presents a device to test flow meters under an arbitrary pressure pulsation in a gas pipeline with a laminar flow containing frequency components up to 50 Hz, with the amplitude reaching hundreds of pascals. In order to reduce flow noise, the device has a strainer-like element connected to a pipeline under test and uses an open-loop control law based on the frequency response test. The control signal is calculated by adding the inputs to obtain each of the sinusoidal waves included in the original wave, which was decomposed by Fourier analysis. The validity of the developed method is demonstrated through the generation tests of superimposed pressure waves containing frequency components up to 50 Hz. Analysis of the relative uncertainty demonstrated the relative uncertainty to be less than 10% when the generated pressure is larger than 360 Pa.

Conflicts of Interest

The authors declare no conflicts of interest.

Cite this paper

Nakao, M. , Kato, T. , Oowaku, T. , Sakuma, H. and Kagawa, T. (2014) Generation of Arbitrary Pressure Pulsation of Wide Frequency Range for Flow Meter Testing in a Laminar Gas Pipeline. Journal of Flow Control, Measurement & Visualization, 2, 125-137. doi: 10.4236/jfcmv.2014.24015.

References

[1] Håkansson, E. and Delsing, J. (1994) Effects of Pulsating Flow on an Ultrasonic Gas Flowmeter. Flow Measurement and Instrumentation, 5, 93-101. http://dx.doi.org/10.1016/0955-5986(94)90042-6
[2] Hanson, B.R. and Schwankl, L.J. (1998) Error Analysis of Flowmeter Measurements. Journal of Irrigation and Drainage Engineering, 124, 248-256.
http://dx.doi.org/10.1061/(ASCE)0733-9437(1998)124:5(248)
[3] Berrebi, J., Martinsson, P.E., Willatzen, M. and Delsing, J. (2004) Ultrasonic Flow Metering Errors Due to Pulsating Flow. Flow Measurement and Instrumentation, 15, 179-185.
http://dx.doi.org/10.1016/j.flowmeasinst.2003.12.003
[4] Durst, F., Heim, U., ünsal, B. and Kullik, G. (2003) Mass Flow Rate Control System for Time Dependent Laminar and Turbulent Flow Investigations.Measurement Science and Technology, 14, 893-902. http://dx.doi.org/10.1088/0957-0233/14/7/301
[5] Kawashima, K. and Kagawa, T. (2003) Unsteady Flow Generator for Gases Using Isothermal Chamber. Measurement, 33, 333-340. http://dx.doi.org/10.1016/S0263-2241(03)00003-4
[6] Funaki, T., Kawashima, K., Yamazaki, S. and Kagawa, T. (2007) Generator of Variable Gas Flows Using an Isothermal Chamber. Measurement Science and Technology, 18, 835-842.
http://dx.doi.org/10.1088/0957-0233/18/3/036
[7] Kato, T., Oowaku, T., Sakuma, H. and Kagawa, T. (2012) Introduction of a Newly Developed Arbitrary Pressure Pulsation Generating Device for Evaluating the Characteristics of Gas Flow Meters and Sensors. JFPS International Journal of Fluid Power System, 5, 16-21.
[8] JIS B2301 (1994) Screwed Type Malleable Cast Iron Pipe Fitting.
[9] JIS B8390 (2000) Pneumatic Fluid Power Components Using Compressible Fluids Determination of Flow Rate Characteristics.
[10] Takatsu, Y. and Masuoka, T. (2002) Production and Dissipation of Turbulence in Porous Media. Transactions of JSME series B, 68, 2601-2605.
[11] Sueki, T., Takaishi, T., Ikeda, M. and Arai, N. (2010) Application of Porous Material to Reduce Aerodynamic Sound from Bluff Bodies. Fluid Dynamics Research, 42, 015004.
http://dx.doi.org/10.1088/0169-5983/42/1/015004
[12] ISO (1995) Guide to the Expression of Uncertainty in Measurement.

  
comments powered by Disqus

Copyright © 2018 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.