Inhibition of the Oversensing of Cardiac Pacemakers in Chest CT

Abstract

Objectives: The purpose of this study is to identify how to manage oversensing of pacemakers in chest CT. Methods: Four different models of pacemakers were examined to select the pacemaker generating oversensing. To the pacemaker with oversensing, intermittent switching X-ray was exposed using ECG-gated CT helical scan system at prospective CTA mode. IVY Model was used to synchronize the ECG. Only during in the alert period that is non-refractory and sensing is available, intermittent switching X-ray (300 msec/sec) was exposed in chest CT. For comparison, the same intermittent switching X-ray (300 msec/sec) was exposed in the refractory period when sensing was not available. Results: Oversensing was detected only in one of the four pacemakers tested. In this pacemaker, oversensing was generated by exposure of the intermittent switching X-ray in the alert (non-refractory) period, but oversensing was not observed in the refractory period. Conclusion: A pacemaker has alert and refractory periods. Oversensing of a pacemaker was found to be inhibited by selective ECG-synchronized exposure in the refractory period. Since all pacemakers have the refractory period, the results of this study can be widely applied to the patients with pacemakers in chest CT, and their chest CT can be operated safely.

Share and Cite:

N. Akiba, M. Takeda, G. Nakaya, O. Nakamura, M. Tsuboi, J. Matsumoto, K. Ito, Y. Okuyama, M. Shimada and K. Yoshikawa, "Inhibition of the Oversensing of Cardiac Pacemakers in Chest CT," Open Journal of Medical Imaging, Vol. 2 No. 4, 2012, pp. 119-124. doi: 10.4236/ojmi.2012.24022.

Conflicts of Interest

The authors declare no conflicts of interest.

References

[1] C. H. McCollough, J. Zhang, A. N. Primak, W. J. Clement and J. R. Buysman, “Effects of CT Irradiation on Implantable Cardiac Rhythm Management Devices,” Radiology, Vol. 243, No. 3, 2007, pp. 766-774. doi:10.1148/radiol.2433060993
[2] O. Nakamura, G. Nakaya, S. Ogashiwa, A. Hashimoto and M. Kondou, “Experimental Study of the Influence on Pacemakers of X-Rays from Angiocardiography Equipment,” Japanese Journal of Radiological Technology, Vol. 64, No. 3, 2008, pp. 335-341. doi:10.6009/jjrt.64.335
[3] N. Umezawa, M. Hirose and T. Shinbo, “Verification of Irradiation Conditions of X-Rays That Influence Implantable Cardiac Pacemakers,” Japanese Journal of Radiological Technology, Vol. 64, No. 7, 2008, pp. 795-804. doi:10.6009/jjrt.64.795
[4] N. Oda, H. Nakajima, H. Abe, S. Koyama and S. Kakeda, “Effect of Diagnostic X-Rays of Implantable Cardiac Pacemakers and Implantable Cardioverter Defibrillators, and It’s Management,” Japanese Journal of Radiological Technology, Vol. 64, No. 7, 2005, pp. 805-813. doi:10.6009/jjrt.64.805
[5] M. Lewis, “Radiation Dose Issues in Multi-Slice CT Scanning,” Impact Technology Update, No. 3, 2005. http://www.impactscan.org/download/msctdose.pdf
[6] S. Yamaji, S. Imai and F. Saito, “Does High-Power Computed Tomography Scanning Equipment Affect the Operation of Pacemakers?” Circulation Journal, Vol. 70, No. 2, 2006, pp. 190-197. doi:10.1253/circj.70.190
[7] T. Ushijima, Y. Yamao, N. Nagasawa, M. Matsuzuki and T. Sasou, “Partial Electrical Reset of CT Irradiation on Implantable Cardiac Devices: Relationship between Reset and Tube Voltage, Tube Current, and Rotation Time,” Japanese Journal of Radiological Technology, Vol. 67, No. 6, 2011, pp. 648-653. doi:10.6009/jjrt.67.648
[8] N. G. Blamires and J. Myatt, “X-Ray Effects on Pacemaker Type Circuits,” Pacing and Clinical Electrophysiology, Vol. 5, No. 2, 1982, pp. 151-155. doi:10.1111/j.1540-8159.1982.tb02206.x
[9] D. L. Hayes, P.J. Wang, D.W. Reynolds, M. Estes, J. L. Griffith, et al., “Interference with Cardiac Pacemakers by Cellular Telephones,” The New England Journal of Medicine, Vol. 336, No. 21, 1997, pp. 1473-1479. doi:10.1056/NEJM199705223362101
[10] M. Niehaus and J. Tebbenjohanns, “Electromagnetic Interference in Patients with Implanted Pacemakers or Cardioverter-Defibrillators,” Heart, Vol. 86, No. 3, 2001, pp. 246-248.
[11] Pharmaceuticals and Medical Devices Safety Information, No. 213, 2005, pp. 3-4. http://www.pmda.go.jp/english/index.html
[12] Pharmaceuticals and Medical Devices Safety Information, No. 221, 2006, pp. 12-15. http://www.pmda.go.jp/english/index.html
[13] Pharmaceuticals and Medical Devices Safety Information, No. 263, 2009, pp. 34-36. http://www.pmda.go.jp/english/index.html
[14] T. Ota, M. Tsuyuki, M. Okumura, T. Sato and T. Kondo, “Evaluation of Exposure Dose Reduction in Multislice CT Coronary Angiography (MS-CTA) with Prospective ECG-Gated Helical Scan,” Proceedings of SPIE, Vol. 6913, 2008.

Copyright © 2024 by authors and Scientific Research Publishing Inc.

Creative Commons License

This work and the related PDF file are licensed under a Creative Commons Attribution 4.0 International License.